The CPX Digital Area Alarm is used to monitor pipeline pressure within a ward, theatre, private rooms, etc.

The pressure is monitored via a transducer in the pipeline downstream from the last AVSU.

This pressure is monitored for both high and low pressure (low only for vacuum).

A fault in the cabling between the pressure transducer and alarm panel would result in a system fault alarm, and as a result, the integrity of the CPX cabling is incomparable.

CPX Digital Area Alarm is available in two different and versatile installation models.

The alarm unit provides an indication of 3 conditions of gas: Normal (green), Low Pressure (red), and High Pressure (red) for up to 6 gas services.

*Vacuum service indicates green for Normal and red for Low Pressure.